OB-Den10D Moving Dental X Ray Unit

OB-Den10D Moving Dental X Ray Unit
Detail

OB-Den10D Moving Dental X Ray Unit

FEATURES:

1 The microprocessor-controlled timer brings high-tech x-ray imaging at your fingertips.

2 The control panel allows an accurate and fast exposure time selection while protecting the head during x-ray examinations.

3 Fitted with 70kv 8ma tube with preheating grid, the machine has been specially designed for intra-oral dental diagnosis applications

4 Complies with local and international regulations on ionizing radiation protection for maximum safety.

5 The soft positioner arm ensures naturally smooth and well-balanced movements. It is easy installed at the height and length.

6 Its ergonomics and design have been studied to meet the requirements of the most semanding practices in matters of floor space optimization.

SPECIFICATION:

Classification: Electro medical, Class I, type B

Power supply:AC220V+10%,50Hz,1KVA

Tube voltage:60kVp

Tube current:8mA

Focus size:1.5mm

Total filtration:2.5mmAL

Exposure time:0.2-4sec

Leak radiation:Outside one meter<0.002mGy/h

Optional:Tube current:0.5mA     Focus size:0.8mm

Size: 148*54*32cm/Carton

Total weight: 57kg

Precautions

1. In order to operate safe, must be installed properly good ground.

2. X-ray machine due to small thermal capacity is limited, should pay attention to intermittent operation is recommended.

3. The machine only during exposure, will produce X-rays, power-off state does not produce X-rays, as long as it does not produce radiation exposure is not under power on state.

4. The machine is in the correct use, any unforeseen circumstances, please immediately cut off the power operator, discontinue use. After the aircraft maintenance standards, it can be used again. Additional filtration plate

5. The machine factory, each X-ray device is equipped with the following specifications: 1.0 / 0.5mm aluminum each one.

6. When disassembling the machine each time, first disassemble the nose and then split the other member.

7. After each use the machine, arm recovery highest point, the position of the head is placed equilibrium position.

8. After each use of medical alcohol exposure in some patients with end beam tube disinfection, cleaning.

9. The weekly machine surface with a dry cloth to clean, inspect each connector contact.

10. Note that the machine maintenance and inspection indicator that shows whether the knobs, switches, remote control is working properly; X-ray source assembly if possible oil leakage; when the X-ray source assemblies exposure, abnormal noise; mechanical portions of the fastening screws loosening the like.

11. The machine rated capacity of 1.1kW, please use the greater than 1.5kW power supply capacity or grid. If the regional grid voltage is not normal, we recommend supporting more than 1.5KW industrial regulator, to prevent damage to the machine voltage is not normal, reducing machine life.